• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. SCEPTER XC OCCLUSION BALLOON (4X11 MM)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. SCEPTER XC OCCLUSION BALLOON (4X11 MM) Back to Search Results
Model Number BC0411XC-ID
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Microvention determined that the issue was confirmed to be isolated to this small lot and was due to a human error.The products are manufactured in tightly controlled environment, the product bioburden level is regularly monitored and kept at very low level.Awareness training has been performed to the personnel involved in the incident.One other scepter xc balloon from the same lot was sent to the same customer (total of three devices).The report has been submitted as mfr report # 2032493-2017-00174.There are no other scepter xc balloons with this lot number that exist in the field.This device was used during the same procedure as reported on mfr.Report # 2032493-2017-00175.
 
Event or Problem Description
It was identified that two scepter xc balloons were inadvertently missed from the terminal sterilization process and shipped to a customer.The physician was immediately notified.Two scepter xc balloons were selected for use on (b)(6) 2017 in a patient that was classified as hh1, clinically in good condition, who was undergoing treatment for a ruptured aneurysm.The first scepter balloon was selected for treatment, but was not used due to the presence of a kink on the shaft.The second scepter balloon was used in the patient to assist with coil placement inside the aneurysm sac (balloon remodeling technique).The decision was made to administer prophylactic treatment with a broad-spectrum antibiotic (tazobac) post-procedure.Currently, the patient does not exhibit signs of infection; there is no elevation of crp or leukocytes.The patient is being closely monitored by the physician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCEPTER XC OCCLUSION BALLOON (4X11 MM)
Common Device Name
BALLOON
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key6803005
Report Number2032493-2017-00176
Device Sequence Number11041736
Product Code MJN
UDI-Device Identifier00842636020591
UDI-Public(01)00842636020591(11)160808(17)190731(10)160808GC3
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K113698
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2017
Device Model NumberBC0411XC-ID
Device Lot Number160808GC3
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2017
Initial Report FDA Received Date08/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-