Catalog Number 5955610
Device Problems Use of Device Problem (1670); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2017
Event Type  malfunction  
Manufacturer Narrative
As reported one of the connectors detached from the balloon while being removed through the trocar.The sample evaluation confirms detachment of one of the connectors from the balloon.Regarding removal of the balloon the ifu prescribes that the user remove the balloon and trocar simultaneously.Based on the information provided and product evaluation the balloon connector appears to have detached from forces applied while removing the balloon through and not simultaneously with the trocar as prescribed in the ifu.Regarding the removal of the balloon the ifu states, "begin removal of the echo ps¿ positioning system by grasping one of the two removal points marked by the dark arrows adjacent to the bard® logo.Begin pulling the positioning system off the mesh in one smooth motion.Continue removing the echo ps¿ positioning system, pulling it up to the tip of the trocar.Remove both the echo ps¿ positioning system and trocar simultaneously.Verify that the echo ps¿ positioning system including the balloon, all mesh connectors, and the inflation tube is fully intact, discard the echo ps¿ positioning system appropriately." no lot number has been provided; therefore a review of the manufacturing records could not be performed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Due to privacy laws in (b)(4), the customer has not provided any patient details.
 
Event Description
It was reported that on (b)(6)2017 a bard ventralight st w/ echo ps was being used during a procedure.As reported during the process of removing the balloon through a 15mm port/trocar one of the connectors detached from the balloon and was retrieved from the patient¿s abdomen.There was no injury to the patient.
 
Manufacturer Narrative
This is an addendum to the initial mdr to document that the lot number has been provided.A review of the manufacturing records was performed and found that the lot was manufactured to specification.To date this is the only reported complaint for this manufacturing lot of 64 units released for distribution on 11/30/2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST W/ ECHO
Type of Device
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
laura sundberg
100 crossings blvd.
warwick, RI 02886
4018258462
MDR Report Key6803875
MDR Text Key83105000
Report Number1213643-2017-00515
Device Sequence Number1
Product Code FTL
UDI-Device Identifier00801741031748
UDI-Public(01)00801741031748
Combination Product (y/n)N
PMA/PMN Number
K122436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2018
Device Catalogue Number5955610
Device Lot NumberHUAX1302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-