|
Findings: pump passed displacement test, rewind test, basic occlusion test, prime test, excessive no delivery test, occlusion test, self test and unexpected alarm error test.No alarm noted.Pump passed all operating currents tested within the spec range and passed off no power test.Unit was programmed with test minilink transmitter and the glucose sensor simulator.Unit communicated properly with glucose sensor simulator and display the 100 value properly on display graph.No lost sensor alarm noted.Pump received with cracked reservoir tube lip, cracked case near display window corners and minor scratches on display window noted.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|