| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a call service alarm (cs 078) issue.There was no indication that this issue lead to an adverse event.This is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 10/11/2017 with the following findings: there were call service (cs 078) alarms observed in the black box.The complaint was replicated during testing.The pump was opened and moisture was observed throughout the pump.
|
| |
|
Search Alerts/Recalls
|