• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/20/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a patient underwent an explant of a tecnis symfony intraocular lens (iol), model zxr00 15.5 diopter due to refractive error from presbyopia.The explanted lens was replaced with a model zlb00 16.5 diopter lens.No additional information provided.
 
Additional Manufacturer Narrative
Additional information received reported that the patient is a male with date of birth (b)(6).Also, the implant date was (b)(6) 2017, the operative eye was the right, and the doctor's name was dr.(b)(6).It was also reported that the correct replacement lens was a zlb00 15.5 diopter.There was no incision enlargement, vitrectomy, or sutures used.Also, there was no patient injury.No additional information was provided.(b)(6).If implanted, give date: (b)(6) 2017.(b)(6) health professional: yes, occupation: physician.All pertinent information available to abbott medical optics has been submitted.
 
Additional Manufacturer Narrative
Device available for evaluation?: yes, returned to manufacturer on 10/05/2017.Device returned to manufacturer?: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key6806231
Report Number9614546-2017-00776
Device Sequence Number19082563
Product Code POE
UDI-Device Identifier05050474591752
UDI-Public(01)05050474591752(17)210908
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/08/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2017
Initial Date Received by Manufacturer 07/26/2017
Supplement Date Received by Manufacturer09/11/2017
10/24/2017
Initial Report FDA Received Date08/19/2017
Supplement Report FDA Received Date10/05/2017
11/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-