| Model Number |
2479 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Multiple attempts for product return and requests for additional information were made.The product has not been returned to masimo to allow an analysis to be performed.If new information is obtained or the product is returned, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported patient is anesthetized enough but psi showed high.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.During evaluation, the sensor passed visual inspection; however, failed continuity testing due to an open connection on r1, r2, l2, cb, and ct electrodes.No functional testing could not be performed since the sensor was returned used and could not be tested on the forehead., corrected data: d1 (brand name) was updated from "sedline" to "sedline sensor" (expiration date) was updated from a blank field to "10/01/2019".(report source) was updated from " foreign; health professional; user facility".To"foreign; company representative; user facility".
|
| |
|