| Model Number |
5201260 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Hole In Material (1293); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer reported an issue encountered with the mps2 console during use.The report stated the console continuously popped the vent valve, and the user would have to manually clamp the vent line.There were no patient complications reported as a result of the alleged issue.The console was returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The console was received and evaluated.The reported complaint was confirmed in the log data.Fluid intrusion was found on the solenoid driver board which was the cause for the vent valve errors.It appeared that the fluid came in from a hole/ tear in the shroud tube that's on the water line underneath the hex block.The shroud was replaced with a new shroud and the solenoid driver board was replaced and the console was reassembled.The root cause of the hole/tear in the shroud tube is unknown.Complaints will be monitored for trending.
|
| |
|