| Model Number |
LSMU0800626 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Receipt of device is not expected.The investigation is currently in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that during treatment, the device was allegedly difficult to load through a 6fr introducer sheath.It was further reported that the health care provider managed and was able to deploy the stent successfully, and removed the delivery system without incident.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that during treatment, the device was allegedly difficult to load through a 6fr introducer sheath.It was further reported that the health care provider managed and was able to deploy the stent successfully, and removed the delivery system without incident.There was no reported patient injury.The lot history records could not be reviewed as the lot number was unknown.The device was not returned for evaluation.Have been reviewed with special attention to the manufacturing and inspection of this product.The device was found to have met specifications prior to shipment.No manufacturing anomalies were identified that may have caused or contributed to the reported event.This is the first reported complaint for this lot number and issue to date.The device was returned for evaluation.The result of the investigation is inconclusive as no sample returned for evaluation.The lot number details were also not provided so the lhr and dhr reviews could not be performed.Based upon the available information a definitive root cause has not been determined.It is possible that patient factors, handling or procedural techniques may have contributed to the reported event.Based on analysis performed no additional action is required at this time.Furthermore while the current confirmed calculated rate for this failure mode is 0.043% (last 24 months) and is hence higher than the predicted rate of 0.02% the rate is decreasing since the disparity check for the msi crimping equipment was introduced into production in jan 17.The disparity check was a corrective action taken through a capa.The current rate from feb to june 17 = 0.017%.This is below the predicted rate of 0.02%.The ifu states: a device description: implant: the lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from 316l stainless steel, encapsulated between two layers of eptfe.Indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use: site access and preparation: using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection: select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation carefully remove the selected device from the package.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation a 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent advance the endovascular system over the guidewire into the introducer sheath.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.(b)(4).
|
| |
|
Search Alerts/Recalls
|