• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4006C0204
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the penumbra coil 400 (pc400) pusher assembly.The embolization coil was intact with the pusher assembly.The px slim delivery microcatheter (px slim)was kinked approximately 2.0 cm from the proximal end.Conclusion: evaluation of the returned devices revealed the pc400 could be advanced through a demonstration microcatheter without issue.Further evaluation revealed the px slim was kinked near the hub.This damage may have occurred due to forceful handling of the px slim during preparation or positioning within patient anatomy.The kinked region likely contributed to the inability to advance the pc400 during the procedure.Penumbra coils and catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2017-01430.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra coil 400's (pc400s).During the procedure, the physician successfully placed pc400s using a px slim delivery microcatheter (px slim).The physician then felt resistance advancing a pc400 through its introducer sheath and into the microcatheter, and was unable to advance the pc400 through the px slim; therefore, the pc400 was removed.It was noticed that the px slim was kinked near the strain relief on the proximal end of the catheter; therefore, the px slim was removed.The procedure was completed using a new pc400 and a new px slim.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6810487
Report Number3005168196-2017-01429
Device Sequence Number985106
Product Code HCG
UDI-Device Identifier00814548011783
UDI-Public00814548011783
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/31/2023
Device Catalogue Number4006C0204
Device Lot NumberF63222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2017
Initial Report FDA Received Date08/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
-
-