| Model Number |
MMT-551PNAB |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Customer called to report that the insulin pump alarmed battery out limit.Customer's blood glucose reading at the time of the incident was not included in the report.No significant events leading to the alarm were observed.Customer was advised to be sent a replacement battery cap.Product is not expected to return.
|
| |
|
Additional Manufacturer Narrative
|
|
Device power up properly after battery installation.No blank display noted.Device was received with normal operating currents and passed self-test, unexpected restart error test.No unexpected off no power, low battery, battery out limit and failed battery test alarms during testing.
|
| |
|