| Model Number |
24055- |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A patient was scheduled for laparoscopic bilateral inguinal hernia repair with mesh placement.The surgeon introduced the auto suture locking trocar and when she attempted to lock the device in place, the "anchors" broke rendering the device inoperable.A second device with the same lot number was used, but resulted in the same defect.A third device was opened, introduced and was successfully deployed, at which point the surgery proceeded as planned.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The visual inspection of the returned products noted; the spiked trocars were received.The circular and envelope seals were intact.The flanges of the anchors were skewed and broken.The condition of the devices precludes functional testing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported condition may be due to insertion or removal of the port without the anchor tabs being fully closed.Under these circumstances, the tabs may be exposed to excessive force and therefore break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|