DAVIS & GECK CARIBE LTD AUTO SUTURE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 24055-
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A patient was scheduled for laparoscopic bilateral inguinal hernia repair with mesh placement.The surgeon introduced the auto suture locking trocar and when she attempted to lock the device in place, the "anchors" broke rendering the device inoperable.A second device with the same lot number was used, but resulted in the same defect.A third device was opened, introduced and was successfully deployed, at which point the surgery proceeded as planned.There was no patient injury reported.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The visual inspection of the returned products noted; the spiked trocars were received.The circular and envelope seals were intact.The flanges of the anchors were skewed and broken.The condition of the devices precludes functional testing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported condition may be due to insertion or removal of the port without the anchor tabs being fully closed.Under these circumstances, the tabs may be exposed to excessive force and therefore break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO SUTURE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key6811495
Report Number9612501-2017-05999
Device Sequence Number2066525
Product Code GCJ
UDI-Device Identifier10884521080829
UDI-Public10884521080829
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K981941
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model Number24055-
Device Catalogue Number24055-
Device Lot NumberJ5H2112X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2017
Initial Date Received by Manufacturer 07/25/2017
Supplement Date Received by Manufacturer10/12/2017
Initial Report FDA Received Date08/22/2017
Supplement Report FDA Received Date11/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
Patient Weight68
-
-