| Model Number |
48759701 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; trial handle tip of knob or trial handle broke off during surgery.2 tabs, one may have already been missing, other may have off.During surgery while inserting the trial, the surgeon softly impacted the handle and the tip of the knob, where the handle connects to the trial, broke off.One tab of the instrument may have already been missing.The other tab fell off during this surgery and was successfully retrieved.While using the second back up handle, when the handle was taken off to take an x-ray, metal fragments came off.There were no adverse consequences to the patient.Surgery was successfully completed with the second trial handle.Both instruments were previously used successfully.Rep had the set for 2 weeks and it was used 3x successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned trial handle was inspected visually.Sign of wear after long-term usage was observed and no fracture can be found.No failure was detected on the returned product.
|
| |
|
Event or Problem Description
|
|
It was reported that; trial handle tip of knob or trial handle broke off during surgery.2 tabs, one may have already been missing, other may have off.During surgery while inserting the trial, the surgeon softly impacted the handle and the tip of the knob, where the handle connects to the trial, broke off.One tab of the instrument may have already been missing.The other tab fell off during this surgery and was successfully retrieved.While using the second back up handle, when the handle was taken off to take an x-ray, metal fragments came off.There were no adverse consequences to the patient.Surgery was successfully completed with the second trial handle.Both instruments were previously used successfully.Rep had the set for 2 weeks and it was used 3x successfully.
|
| |
|
Search Alerts/Recalls
|