| Model Number |
MMT-522NAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump alarmed motor position encoder error.The customer¿s blood glucose level was unknown at the time of the incident.The customer also reported to have received a motor error alarm and during troubleshooting the customer stated that the drive support cap was normal and that the insulin pump was not exposed to high magnetic fields.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the rewind, basic occlusion, prime and displacement test.Insulin pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.Unable to verify motor position encoder error alarm in the alarm history due to history file overwritten.
|
| |
|