| Model Number |
4001-9 |
| Medical Device Problem Code |
Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned sensor was evaluated.The unit passed visual inspection and software testing; however the device did not pass functional testing.The sensor was failed continuity due to an open emitter assembly.During testing, the sensor was able to generate a "replace sensor" error message.
|
| |
|
Event or Problem Description
|
|
The customer reported the new sensors has a high failure rate out of the box and intermittent failure afterwards, the red light would not light up or would not stay on.No patient impact or consequences were reported.
|
| |
|