| Model Number |
VLOCN006L |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a lap nissen, while suturing the gastric wrap together, the thread of the device broke.Another device was used to complete the procedure.No patient injury noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: product analysis was created by the (b)(4) pmv lab as part of the receipt process, so that decontamination could be documented in (b)(6).This product analysis activity is being marked as ¿complete¿ and the investigating site will create another product analysis activity upon receipt.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted one suture and one needle.The loop was received open, and it was observed, under magnification, that it appeared to have broken under tension.Under microscopic inspection, evidence of material transfer was observed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|