Catalog Number 190610
Device Problems Charred (1086); Smoking (1585)
Patient Problem No Patient Involvement (2645)
Event Date 07/13/2017
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician (biomed) at a user facility reported to technical support that a 2008k2 hemodialysis (hd) machine had a flow error message in heat disinfect.The biomed replaced float switch to resolve the issue.Subsequently, the machine had an acid pump no eos message in dialysis mode.The biomed found a pinched cable on the acid pump, replaced the acid pump and the actuator/test board to resolve the issue.Follow-up indicated that the actuator test board was charred.Smoke was also noted during the repair.Subsequently, the machine is experiencing high conductivity and water coming from the vent tube in heat disinfect.The current repair status of the machine is unknown and it is not known if the unit has been returned to service at the user facility.
 
Manufacturer Narrative
Plant investigation: the device was not returned to the manufacturer for physical evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances or any associated rework during the manufacturing process which could be associated with the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6814333
MDR Text Key83532288
Report Number2937457-2017-00769
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100859
UDI-Public00840861100859
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Biomedical Engineer
Device Catalogue Number190610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Initial Date Manufacturer Received 07/31/2017
Initial Date FDA Received08/23/2017
Supplement Dates Manufacturer Received09/14/2017
Supplement Dates FDA Received10/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-