| Model Number |
2200P-47E |
| Medical Device Problem Codes |
High impedance (1291); Device Inoperable (1663); Device Displays Incorrect Message (2591); Therapeutic or Diagnostic Output Failure (3023)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This patient experienced a flashing red light on their device approximately (b)(6) 2017.On (b)(6) 2017, the site reported that they could not restart therapy.The clinician programmer indicated error code 25 (too many therapy retries).Impedance testing indicated high posterior tip to ring and anterior tip to posterior tip impedance measurements.Device remains implanted, but therapy is not being delivered.
|
| |
|
Additional Manufacturer Narrative
|
|
Engineering was unable to confirm the customers complaint as the rnr passed the final functional test and due to the condition of the leads after removal surgery, they were not able to be investigated.
|
| |
|
Event or Problem Description
|
|
This patient experienced a flashing red light on their device approximately (b)(6), 2017.On (b)(6) 2017, the site reported that they could not restart therapy.The clinician programmer indicated error code 25 (too many therapy retries).Impedance testing indicated high posterior tip to ring and anterior tip to posterior tip impedance measurements.
|
| |
|