• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R Back to Search Results
Model Number 5DCS
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection, the device showed signs of clinical use.Debris was present on the blades of the device.The device was inspected under magnification which revealed wear on the blades.Functional inspection revealed the blades were dull and failed to provide a clean cut.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root cause(s) are: - user attempts to cut staples, clips, or non-soft tissue - wear to blades due to clinical use the instructions for use (ifu) state: - do not directly apply current to staples or clips.- instruments were designed for cutting soft tissue.Attempting to cut staples or clips may damage the instrument.- if cutting of staples or clips is attempted, damage to instrument may occur.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported the laparoscopic scissor is defective.The complainant is unaware of any patient injury, medical intervention, or extended procedure time.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
MDR Report Key6815879
Report Number0001056128-2017-00111
Device Sequence Number16390091
Product Code NUJ
UDI-Device Identifier00885825013288
UDI-Public(01)00885825013288(17)190525(10)46068858
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K110189
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2019
Device Model Number5DCS
Device Catalogue Number5DCSRR
Device Lot Number4606858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2017
Initial Report FDA Received Date08/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-