| Model Number |
H7493808015320 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was tightly folded.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 69.5cm from the hub.The fracture faces were oval as if kinked prior to separation.There was no evidence of any material or manufacturing deficiencies contributing to the damage.Inspection of the remainder of the device presented no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2017.It was reported that crossing difficulties were encountered.The target lesion was located in a coronary vessel.A 3.25mm x 15mm quantum¿ maverick¿ balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed a hypotube break.
|
| |
|