| Model Number |
D902 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient identifier was not provided.Patient sex was not provided.Patient weight was not provided.The age of the device was calculated as the time elapsed between device sterilization and the date of the event.(b)(4).(b)(4) manufactures the d902 dideco lilliput 2 infant hollow fiber oxygenator.The incident occurred in (b)(6).Per exemption number e2016005.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.No other similar complaints have been recorded on this device in the past three years.The involved device is not available for return to sorin group (b)(4) for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not available to be returned.
|
| |
|
Event or Problem Description
|
|
Sorin group (b)(4) received a report that a blood to water leakage occurred in a lilliput 2 infant hollow fiber oxygenator during a procedure.A new oxygenator could not be primed due to the unavailability of blood so the patient temperature was recovered using a heater cooler and blanket.This reportedly took 2 hours.There was no report of patient injury.
|
| |
|