| Model Number |
SP-5698 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, on (b)(6) 2017, pre-operatively, the device broke, there was a damage at the level of the eye of the needle.No patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned samples noted one partial suture and one needle.The retainer was inspected with no abnormalities.The needle was broken/damaged at the swage end and detached from suture.It was observed, under magnification, that instrument marks were present on the swaged end of the needle.As no sealed products were returned, functional testing was precluded.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.The observed instrumentation damages suggested that the needles were grasped improperly at the swaged end of the needle during application.Firmly grasping the needle at the swaged end can deform the crimp, can cause the needle to disengage from the suture, and can cause the needle to break at the swaged end.The recommended grasping area of the needle is at the needle body, where the wire is of a full diameter and significantly stronger than at the swaged end.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|