| Model Number |
00711553 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history record indicates no issues during manufacture; the device was manufactured to specification.The subject device was not returned to us endoscopy.The instructions for use contain the following: "prime the air/water channel prior to insertion of the endoscope into a patient.If water pressure is low, check to see that the lid cap is tightly closed.Check that the endoscope connector is firmly attached to the endoscope." in-service training has been provided to the user.
|
| |
|
Event or Problem Description
|
|
The aquashield system co2 - fujinon is intended to be used with an air or carbon dioxide (co2) source from an endoscope with the purpose of supplying sterile water to the endoscope during endoscopic procedures.It is compatible with u.S.Commercially available sterile water bottles.This device is intended to be used for not more than 24 hours.The user facility reported during a procedure the connection between the aquashield system co2 and the endoscope was leaking sterile water.There was no report of injury to the patient or the user.
|
| |
|