• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0998-00-0800-53
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Failure to Calibrate (2440)
Patient Problem No Information (3190)
Event Date 08/02/2017
Event Type  Injury  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformances related to the reported event were noted.It was determined that the iabp did not malfunction.
 
Event Description
The customer reported while the patient was in a code situation in ccl the fiber optic catheter failed to calibrate and they were unable to obtain hemo tracing.Resorted to fluid transducer pressure from central linen.During swap out of main cable it was noted that the tracing had been frozen by accident and now is working fine.The doctor states smooth uneventful insertion.Patient transferred successfully.The patient coding is not attributed to the iabp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key6817579
MDR Text Key83519131
Report Number2249723-2017-00336
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeYR
Date Manufacturer Received08/02/2017
Date Device Manufactured07/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age60 YR
-
-