• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 HYDROSIL GO¿ INTERMITTENT URINARY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 HYDROSIL GO¿ INTERMITTENT URINARY CATHETER Back to Search Results
Model Number 71812
Health Effect - Clinical Code Discomfort (2330)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheters were not smooth.The nurse used them with a patient, to teach him intermittent catheterisation.The catheters were rough and uncomfortable for the patient.There was no medical intervention.
 
Additional Manufacturer Narrative
The reported event could not be confirmed.While the samples passed the visual evaluation, a lubricity functional evaluation could not be performed since the samples were inadvertently left open/exposed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿review the self-catheterization procedure with a healthcare professional.Always wash hands prior to use.Open the catheter package by peeling the tab upwards with the aid of the finger hole.If desired, hang the package with the adhesive surface on the inner side of the tab to a nearby dry vertical surface while preparing to catheterize.Positioned comfortably with thighs spread apart, clean the opening to the urethra ¿ and the surrounding area.Wipe from front to back to avoid fecal contamination.Wash your hands again.Remove the catheter with the aid of the insertion handle and advance the catheter tip into the urethra.Slowly and gently insert the catheter into the urethra until urine begins to flow (approximately 1-1.5" or 2.5-3.8 cm).When urine stops flowing, begin to withdraw the catheter.It is recommended to slowly rotate the catheter during withdrawal, stopping each time urine begins to flow.Check the color, odor and clarity of the urine.Any changes may need to be reported to a health care professional.Disposal place the catheter back into the package and dispose in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Do not flush down the toilet.Wash hands." (b)(4).
 
Event or Problem Description
It was reported that the catheters were not smooth.The nurse used them with a patient, to teach him intermittent catheterisation.The catheters were rough and uncomfortable for the patient.There was no medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSIL GO¿ INTERMITTENT URINARY CATHETER
Common Device Name
HYDROSIL
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6817777
Report Number1018233-2017-04420
Device Sequence Number1443910
Product Code EZD
UDI-Device Identifier00801741097997
UDI-Public(01)00801741097997
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K150345
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2018
Device Model Number71812
Device Catalogue Number71812
Device Lot NumberJUAT0640
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2017
Initial Date Received by Manufacturer 08/03/2017
Supplement Date Received by Manufacturer11/29/2017
Initial Report FDA Received Date08/24/2017
Supplement Report FDA Received Date11/30/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-