COOK ENDOSCOPY FUSION LITHOTRIPSY EXTRACTION BASKET; FFL, DISLODGER, STONE, BASKET, URETHRAL, METAL
|
Back to Search Results |
|
| Catalog Number |
FS-LXB-3X6 |
| Medical Device Problem Code |
Split (2537)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: acrobat calibrated tip wire guide (acro-35-260), fusion wire guide locking device (fs-wl-o-s).Investigation evaluation: our evaluation of the product said to be involved confirmed the report.The orange wire guide lumen on the distal end of the device has been stripped along its full length.The basket was returned partially retracted into the sheath.The basket extends and retracts as intended.The catheter has a kink 14 cm from the distal tip of the catheter.The device was returned with an unknown reddish liquid on the outside of the device, in the catheter lumen, and on the basket.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use state the following: "introduce device into endoscope accessory channel, relock wire guide into wire guide locking device.With elevator open, advance basket in short increments until endoscopically visualized exiting scope." prior to distribution, all fusion lithotripsy extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion lithotripsy extraction basket.As reported to customer relations: "the basket was advanced over the guide wire and down the endoscope.When the basket began to exit the endoscope, the physician noticed the angle of the basket was incorrect.The basket was then removed and inspected.It was then discovered that the guide wire had ripped through the orange sheath.".
|
| |
|
Search Alerts/Recalls
|
|
|