MAUDE Adverse Event Report: AMEDA, INC. PURELY YOURS ULTRA; ELECTRIC BREAST PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMEDA, INC. PURELY YOURS ULTRA; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24501879
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/28/2017
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and did not meet functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment and pump.Internally, no signs of foreign substance, burning, melting or charring were observed.The returned batteries were observably damaged.The battery fluid crust was observed on returned batteries.Additional testing confirmed that batteries may leak when the upper middle battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2017 to report hearing strange sizzling sounds from inside the purely yours ultra pump base while pumping on (b)(6) 2017.She was using batteries to power on her breast pump at the time.She stopped pumping, opened the door to the battery compartment and noted nothing unusual so resumed pumping.After pumping was completed, she looked inside compartment again and noted leaking black fluid coming from the batteries.Customer denies coming in contact with the leaking battery fluid therefore no injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS ULTRA
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6818155
MDR Text Key83626707
Report Number3009974348-2017-00285
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 07/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24501879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age31 YR
-
-