Model Number VLOCM0013
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the device broke, the loop at the end of the device kept breaking.No patient injury.The device is expected to be returned for evaluation.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted one suture and one needle.The suture was received with an open loop.It was observed, under magnification, that the suture appeared to have split under tension.Evidence of material transfer was also observed under microscopic inspection of the loop.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC 90
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
COVIDIEN NORTH HAVEN - MFG
195 mcdermott road
north haven CT 06473
MDR Report Key6819545
Report Number1219930-2017-06685
Device Sequence Number121493
Product Code GAM
UDI-Device Identifier20884521131641
UDI-Public20884521131641
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K100257
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Model NumberVLOCM0013
Device Catalogue NumberVLOCM0013
Device Lot NumberA7E0114VFX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2017
Initial Date Received by Manufacturer 07/26/2017
Supplement Date Received by Manufacturer12/21/2017
Initial Report FDA Received Date08/24/2017
Supplement Report FDA Received Date01/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-