|
Conclusion: as reported by a healthcare professional, a kink was found on the fore-end of the deltapaq coil (cdf10015430/c20524).They used another one to complete the procedure.There was no report of patient injury.It was reported that the device would be returned for analysis.Multiple attempts to obtain additional information were unsuccessful.The device was returned for analysis.There is a small length of embolic coil exposed from the end of the green introducer.There is a severe kink in the dpu core wire approximately 36 cm from the distal end of the device; the dpu core wire protrudes from the translucent introducer sheath at the kink.There is a slight bend in the dpu core wire approximately 38 cm from the proximal end of the device.The embolic coil was retracted into the translucent introducer, and the length was measured with ruler.The extended coil is 4.2 cm long, which is within the specification range of 4.0 ¿ 4.5 cm for this item number.The ball tip is absent.There are no kinks visible on the embolic coil.The v-notch is undamaged.The embolic coil was advanced out of the green introducer.No kinks or stretched sections of embolic coil can be seen.The rh coil has not received heat and melted.The v-notch is undamaged.A review of manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection processes related to the reported complaint.The complaint that the embolic coil is kinked is not confirmed.There are no kinked or stretched sections of embolic coil present.However, the evidence of the absent ball tip suggests that there has been damage to the distal end of the embolic coil, which may have presented as a kink and then later have broken off.The severe kink in the dpu core wire is additional evidence of application of excessive force, which may have resulted in the reported issue.The exact sequence of events surrounding the reported incident are not known, so it cannot be determined whether the damage was present before the device was used, or if it became damaged during use.The presence of the ball tip and condition of the embolic coil are verified 100% in-process, and are further inspected in-process at special inspection level.Thus, it is very unlikely that the device left the manufacturing facility with the ball tip missing or kinked.There is no current safety signal identified related to the reported event based on review of complaint history for the device.Since there was no evidence to suggest the event was related to a manufacturing issue, no corrective actions will be taken at this time.
|