DEPUY SYNTHES POWER TOOLS OSCILLATING SAW ATCH, LARGE, WITH KEY; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 532.026VET
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported that during an unspecified veterinary surgical procedure, it was observed that the oscillating saw attachment device just stopped working.It was reported that there was an hour delay in the procedure due to the event as an unspecified spare device was available for use to complete the procedure successfully.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
The date returned to manufacturer was documented as aug 14, 2017 on the initial report.This date has been updated to aug 30, 2017.Device evaluation: the device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device was evaluated and it was determined that bearing of the attachment device had seized, jammed, and was moving heavily.It was noted that the bearing was blocked.It was further determined that the device failed pretest for check frequency.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILLATING SAW ATCH, LARGE, WITH KEY
Type of Device
BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6820236
MDR Text Key83641481
Report Number8030965-2017-14604
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.026VET
Device Lot Number100147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received09/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-