DEPUY SYNTHES POWER TOOLS OSCILLATING SAW ATCH, LARGE, WITH KEY; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 532.026VET |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported that during an unspecified veterinary surgical procedure, it was observed that the oscillating saw attachment device just stopped working.It was reported that there was an hour delay in the procedure due to the event as an unspecified spare device was available for use to complete the procedure successfully.There was no human patient involvement as this was a veterinary procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
The date returned to manufacturer was documented as aug 14, 2017 on the initial report.This date has been updated to aug 30, 2017.Device evaluation: the device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device was evaluated and it was determined that bearing of the attachment device had seized, jammed, and was moving heavily.It was noted that the bearing was blocked.It was further determined that the device failed pretest for check frequency.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|