| Model Number |
RELTACK8RDPT |
| Medical Device Problem Code |
Failure To Adhere Or Bond (1031)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, while tacking in mesh during the laparoscopic umbilical hernia procedure, the device was not able to penetrate and hold correctly onto the tissue of the patient.They opened another tacker to complete the case.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The reload was received partially fired and with disrupted timing.Microscopic inspection showed no damage to the inner tube.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures; therefore, a device history record will not be performed.Replication of the reported condition could be due to excessive manipulation during use.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|