| Model Number |
3L79-31 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, review of instrument logs and a review of labeling.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely depressed calcium results on the architect c16000 analyzer.The following data was provided: patient 1 initial 3.3, repeat 9.4 mg/dl; patient 2 initial 7.8, repeat 9.1 mg/dl ; patient 3 initial 5.6, repeat 9.1mg/dl; there was no impact to patient management reported.
|
| |
|
Search Alerts/Recalls
|