| Model Number |
48394005 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; this device was used at an extreme angle under exceeding pressure.Update 8/2/2017: the tip broke.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported event was confirmed via functional inspection.The internal mechanism of the device was found broken.Manufacturing history was reviewed and no issues were identified.A root cause of the event was determined to be patients' tight disk space as well as using the device at extreme angle.
|
| |
|
Event or Problem Description
|
|
It was reported that; this device was used at an extreme angle under exceeding pressure.Update 8/2/2017: the tip broke.
|
| |
|