| Model Number |
ZXT300 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.Attempts have been made to obtain missing information; however, to date, no response has been received.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis symfony toric intraocular lens (iol), model zxt300 20.0 diopter, was explanted from the patient's right eye due to ''breakdown of the iol''.There was no wound enlargement, complications, or sutures used.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation?: yes, returned to manufacturer on 09/13/2017.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|