|
Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.During the evaluation of the returned device, the device was functionally tested.When the handle of the device was manipulated the forceps cups will open, but they will not close.Several attempts were made to close the forceps cups, but they were unsuccessful.The device was then placed down an olympus gif-q20 (2.8 mm channel) endoscope.The endoscope was put in a torturous path.When the handle was manipulated the device would open, but the cups would not close.The device was sent back to the supplier for evaluation.The supplier provided the following evaluation: one device from the reported event was returned in a zip type bag with proof of decontamination.Evaluation of the returned device was performed.The device was tested for "would not open or close".During functional testing, with the device coiled in two (2), 8" loops, it was confirmed that the device would not operate properly when the handle was manipulated.The reported defect was confirmed; however, the device does open but will not close.Upon further investigation and disassembly of the device tip, it was noted that the device has a butt joint that has broken at the solder connection.Failure was due to a broken solder joint.The device history records for two (2) packaging work orders were reviewed.The first packaging work order consisted of one assembly order, manufactured february 2017.The second packaging work order consisted of one assembly order, manufactured november 2016.Relevant defects were noted in the manufacturing records and the final quality control checklist.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the supplier provided: the 'forceps would not close' issue experienced by the customer was confirmed and the root cause was determined to be a broken solder joint.The supplier is currently working on improvements to create a more robust soldering process to prevent solder joints from breaking.In addition, a corrective action has been initiated to reduce occurrences of cups open/close difficulties for disposable forceps.The product said to be involved is included in the scope of the corrective actions.The instructions for use state the following in regard to product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forceps is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura biopsy forceps are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
In preparation for a procedure, the user selected a cook captura serrated forceps with spike.The cups did not open or close smoothly when the customer checked device function prior to patient contact.Another captura serrated forceps with spike was used instead for the procedure.The device was evaluated on 08/08/2017.It was found that the cups would open, but they would not close at all.
|