• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC. FREESTYLE FREEDOM; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC. FREESTYLE FREEDOM; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 11001-02
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.Note: the meter is designed to report readings of 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings less than 20 mg/dl.A "lo" display message indicates a reading less than 20 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 140 mg/dl and 19 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The reported products were returned and investigated.The complaint was not confirmed and no new issues were observed.All results were within range specification and no errors were observed during control solution testing.The reported readings were not found in the meter's memory.Product was found to be performing within specification.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc blood glucose meter within 10 minutes.Results of 140 mg/dl and 19 mg/dl were plotted on the parkes error grid.The results fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE FREEDOM
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC.
1360 south loop road
alameda CA 94502 7001
MDR Report Key6827702
Report Number2954323-2017-06005
Device Sequence Number19229840
Product Code NBW
UDI-Device Identifier00699073707925
UDI-Public(01)00699073707925(10)1631302(17)181130
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K092638
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2018
Device Model Number11001-02
Device Catalogue Number70792
Device Lot Number1631302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Initial Date Received by Manufacturer 08/11/2017
Supplement Date Received by Manufacturer12/11/2017
Initial Report FDA Received Date08/29/2017
Supplement Report FDA Received Date12/19/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight77
-
-