• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR) Back to Search Results
Model Number 125039
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/04/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a communication that during the cuff test, the air from the tracheal balloon was not deflated completely.The customer indicated that there was no patient involvement associated to this event.
 
Additional Manufacturer Narrative
The sample was returned and visual examination showed that the shape returned unit had been altered using the stylet wire.The stylet wire was removed and inflation/deflation testing was successful.There was no fault found with the device.The probable root cause is that the cuff test was carried out when the shape of the tube was incorrectly altered with the stylet wire.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a communication that during the cuff test, the air from the tracheal balloon was not deflated completely.The customer indicated that there was no patient involvement associated to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, BRONCHIAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key6827771
Report Number8020889-2017-05127
Device Sequence Number64968
Product Code BTS
UDI-Device Identifier30884522008758
UDI-Public30884522008758
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K771219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2021
Device Model Number125039
Device Catalogue Number125039
Device Lot Number201601005X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2017
Initial Date Received by Manufacturer 08/07/2017
Supplement Date Received by Manufacturer11/03/2017
Initial Report FDA Received Date08/29/2017
Supplement Report FDA Received Date11/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-