| Catalog Number |
456003 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device involved was confirmed not to be available for evaluation.Without the actual device, a thorough investigation could not be performed and the reported defect was not confirmed.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If a sample becomes available, a follow up will be submitted.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: the stopcock leaked blood during use.While returning blood to patient through a closed system, approximately 2-3 cc of blood leaked out of the distal stopcock connection onto the patient and the patient's mother.The patient's closed system was then flushed with 10 cc of normal saline and 300 units of heparin to the patient's port and then the patient was de-accessed.No patient injury reported.
|
| |
|