Findings: pump passed displacement test, rewind test, basic occlusion test, prime test, excessive no delivery test, occlusion test, self test and unexpected alarm error test.Pump passed all operating currents tested within the spec range and passed off no power test.Pump was monitored and tested with no blank display noted.Pump received with corroded battery tube, cracked reservoir tube lip, cracked case near display window corners and minor scratches on display window noted.The battery cap was inspected and no anomaly was noted.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|