| Catalog Number |
AR-8967-1855 |
| Health Effect - Clinical Codes |
Erythema (1840); Pain (1994); Swelling (2091); Toxicity (2333)
|
| Date of Event |
03/31/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.This is one of two submissions from the same case.The other is (b)(4).Lot number was not provided so device history record review cannot be performed.The contribution of the device to the reported event could not be determined as the device remained in the patient therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.A possible cause of this type of event may be a reaction of the patient to the material(s) implanted or used during the implant procedure.Product directions for use warns of effects like foreign body and allergic-like reactions as well as infections both deep and superficial to the implant materials.Patient sensitivity to materials must be considered prior to implantation.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an a left ankle fusion procedure on (b)(6) 2013 for a broken ankle and has since underwent numerous surgeries as she fell and broke the same ankle again during recovery.The hardware from the previous surgeries were all from a different manufacturer.On (b)(6) 2016, she underwent an arthroscopic ankle fusion procedure.During this surgery, the hardware previously implanted was removed and the following screws were implanted.Ar-8967-1855, lot: unknown, ar-8967-1845, lot: unknown.Since that surgery, the patient has been experiencing swelling, redness, warmth and radiating pain in her left foot.She has also experienced fatigue.The patient also went to see an allergist which confirmed that she was indeed allergic to nickel.Her surgeon is recommending a revision surgery, however, the patient is hesitant, as this will be her 7th surgery.She is looking to see if the devices listed above contain nickel.
|
| |
|
Search Alerts/Recalls
|
|
|