• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS SOFT-VU SIZING ANGIOGRAPHIC CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS SOFT-VU SIZING ANGIOGRAPHIC CATHETER Back to Search Results
Catalog Number H787107093015
Medical Device Problem Codes Crack (1135); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.The firm is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on (b)(6) 2017: during prep for an angiographic procedure, when the catheter was removed form the sterile packaging, it was noted the tip of the catheter had detached inside of the packaging.The device was set aside and a new of the same was used to complete the procedure.It was reported the defective disposable device is available for return to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was one soft-vu catheter.A visual review of the catheter noted that the tip was fractured,.4cm from the weld joint.The customer's reported complaint description of a tip fracture is confirmed.Although the reported complaint description is confirmed, a root cause for the event cannot be determined.Capa (b)(4) was initiated to determine the root cause and implement a corrective / preventative action for this type failure.A review of the lot history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number, contains a statement; "reshaping of catheter tip is not recommended.Physical damage to the catheter material can result when exposed to heat".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFT-VU SIZING ANGIOGRAPHIC CATHETER
Common Device Name
SOFT-VU SIZING ANGIOGRAPHIC CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
MDR Report Key6833566
Report Number1319211-2017-00086
Device Sequence Number18848244
Product Code DQO
UDI-Device IdentifierH787107093015
UDI-PublicH787107093015
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K112452
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date09/30/2017
Device Catalogue NumberH787107093015
Device Lot Number4813871
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2017
Supplement Date Received by Manufacturer08/18/2017
Initial Report FDA Received Date08/31/2017
Supplement Report FDA Received Date10/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-