| Catalog Number |
H787107120065 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
08/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported to angiodynamics on august 14, 2017: it was reported that during an angiographic procedure, the tip of the soft-vu sos 2 detached inside of the patient.The treating physician successfully removed the fractured tip via snare.The device was set aside and a new of the same device was used to complete the procedure.The reported defective disposable device has been returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned for evaluation was a soft-vu catheter.A visual review of the catheter noted that the tip is fractured.The fracture occurs 5mm from the weld joint, on the soft tip.The customer's reported complaint description of soft-vu tip fracture is confirmed.Although the reported complaint description of catheter tip fracture is confirmed, a definitive root cause cannot be determined.A review of the lot history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Capa (b)(4) was initiated to determine the root cause and implement a corrective / preventative action for this type failure.The instructions for use, which is supplied to the end user with this catalog number, contains a statement; "reshaping of catheter tip is not recommended.Physical damage to the catheter material can result when exposed to heat".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
|
| |
|
Search Alerts/Recalls
|