Model Number Distal Femur Plate |
Device Problems
Failure To Adhere Or Bond (1031); Manufacturing, Packaging or Shipping Problem (2975)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/21/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Examination of the involved plate revealed that the recess intended for handle connection was located with a 0.1 mm offset; this results in a difficult (although possible) plate-handle connection.Following investigation of the case, it was concluded that the reported case was a result of a manufacturing fault.This report is submitted following an fda inspection at the manufacturer facility.
|
|
Event Description
|
During operation, it was difficult to connect a piccolo composite distal femur plate and its handle (outside the patient body).Another plate was used.
|
|