• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT225
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zct225 17.0 diopter was explanted due to an incorrect lens being placed into the operative eye.There was no medical injury reported.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation?: yes, returned to manufacturer on 08/31/2017.Device returned to manufacturer?: yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye showed the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key6837075
Report Number9614546-2017-00825
Device Sequence Number18847197
Product Code HQL
UDI-Device Identifier05050474552678
UDI-Public(01)05050474552678(17)210329
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date03/29/2021
Device Model NumberZCT225
Device Catalogue NumberZCT225U170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2017
Initial Date Received by Manufacturer 08/08/2017
Supplement Date Received by Manufacturer09/25/2017
Initial Report FDA Received Date08/31/2017
Supplement Report FDA Received Date10/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-