MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problem Misconnection (1399)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/09/2017
Event Type  Injury  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose.Once the evaluation will be performed, a follow-up medwatch report will be submitted.
 
Event Description
It was reported that during the procedure pre-surgery it was impossible to connect the robot.The surgeon decided to abort the computer-assisted surgery and perfom a non robotic surgery.
 
Manufacturer Narrative
The device (b)(4) has been inspected for investigation purpose.According to results of the investigation, based on functional inspection, there was a database corruption that may be due to a device reboot after an improper shut down.The improper shut down root could not be determined, it may be due to a power loss or a system crash.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key6837337
MDR Text Key84309860
Report Number3009185973-2017-00700
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberROSA 2.5.8
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-