| Model Number |
MT22430-PNK |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4)2017 that on (b)(6)2017 the receiver had no audio output.No additional event or patient information is available.No product or data were provided for evaluation.The reported no audio output could not be confirmed.A root cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint device was returned for evaluation.The device was externally visually inspected and no defect was found.Receiver was charged and booted up.Functional test was performed and failed.Manual "try it" test was performed and failed.Performed manual test by setting sound level to 55 and no audio was heard.Global receiver communication tool test was performed and failed.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker assembly.
|
| |
|