Model Number 404011
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Swelling (2091); Numbness (2415)
Date of Event 04/27/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load implant failure.This implant loss suggests that the source of the problem was likely to stem from procedural errors and/or the lack of osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, or patient behavior.Proper intended use of the implant is described in its instructions for use.Physical analysis was not performed.
 
Event or Problem Description
Pain, numbness, swelling and fistula were reported.Bone quality at the time of implant failure type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability was achieved.Mobility was also reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTERNAL HEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key6839397
Report Number1287163-2017-51044
Device Sequence Number1420065
Product Code DZE
UDI-Device Identifier00614950005475
UDI-Public00614950005475
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K954513
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Model Number404011
Device Catalogue Number404011
Device Lot Number15060043
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2017
Initial Report FDA Received Date09/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-