| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/31/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician used an evercoss pta balloon with a 7fr sheath and an unknown guidewire for the treatment of an unknown lesion type in the iliac vein (diameter 12mm).Embolic protection was not used.The ifu was followed and device was prepped without issue.The balloon was inflated using an unknown inflation device with a contrast/saline fluid to a pressure of 8 atm.It was reported that a break occurred at the sheath during removal.The detached portion remains in the patient.It is unknown if resistance was encountered or if excessive force was used.In order to complete the procedure, the remaining balloon fragment was stented over within the vessel (unknown size stent used).No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: sanguine residue was noted within the inflation lumen port of the y-manifold of the returned evercross dilatation catheter.The proximal end of the balloon chamber exhibited radial and longitudinal tearing.The inner guidewire lumen exhibited a separation face at the distal end of the balloon chamber¿s proximal marker band.The inner guidewire lumen exhibited longitudinal stretching just proximal of the proximal edge of the proximal radiopaque marker band, (the distance between the parallel light reflections narrows down).The distal edge of the radiopaque marker exhibits an oval shape.Only approximately 76.3cm of the catheter distal of the y-manifold¿s printed strain relief was returned for evaluation.Over 40mm of balloon chamber length and inner guidewire lumen were not returned.The distal marker band and distal tip were not returned.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|