Model Number 343-10-707
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/04/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a wash out and poly exchange.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the surgeon performed a washout and poly exchange.The in-vivo length of patient service for the implant was 21 days.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.The complaint states the surgeon did a wash out and a poly exchange for the patient.No other reason or cause was offered for the procedure.It is unknown if this procedure addressed a possible infection.No information was submitted with the complaint regarding pre-existing conditions that may suggest an infection or inhibited the patient's immune system.This event is deemed as non-product related.The root cause for this event was the surgeon performed a washout and poly exchange.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR PS KNEE
Common Device Name
EMPOWR PS KNEETM TIBIAL INSERT, SIZE 7, 10MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6840421
Report Number1644408-2017-00724
Device Sequence Number706124
Product Code JWH
UDI-Device Identifier00888912171991
UDI-Public(01)00888912171991
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K160342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/27/2021
Device Model Number343-10-707
Device Catalogue Number343-10-707
Device Lot Number011U1005
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/04/2017
Supplement Date Received by Manufacturer09/21/2017
Initial Report FDA Received Date09/01/2017
Supplement Report FDA Received Date10/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-