• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977C290
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Neurological Deficit/Dysfunction (1982); Stenosis (2263)
Date of Event 09/01/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received via a manufacturing representative (rep) regarding a patient who was implanted with a neurostimulator (ins).Information was reported that during a procedure as provider was trying to center lead in epidural canal patient lost neurological function on right side of body as reported by neuromonitoring representative.The surgeon and physician assistant immediately removed the lead and began to flush the area repeatedly.The patient gained full motion again, but provider decided to cancel the case and not proceed.The patient has a stent in his brain and other unknown issues.No further complications were reported.No additional patient symptoms were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received.The healthcare provider said the cause of the patient losing neurological function was likely related to mild cervical stenosis.No further complications reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFY SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key6841886
Report Number2649622-2017-10154
Device Sequence Number18962504
Product Code GZB
UDI-Device Identifier00643169464667
UDI-Public00643169464667
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/30/2020
Device Model Number977C290
Device Catalogue Number977C290
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2017
Supplement Date Received by Manufacturer10/20/2017
01/14/2019
Initial Report FDA Received Date09/05/2017
Supplement Report FDA Received Date10/24/2017
01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-