MAUDE Adverse Event Report: DATASCOPE FAIRFIELD UNKNOWN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD UNKNOWN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number UNKNOWN
Device Problem Unraveled Material (1664)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.(b)(4).
 
Event Description
Customer states that their ccl is having difficulty with insertion of intra-aortic balloon (iab) catheters.They inserted the sheath but was not able to pass the iab through the sheath.The membrane starts to unravel and cannot be passed through the sheath.According to the customer this has happened in the past but no catheters were saved.The customer did not provide any specific product, event, or patient information.There has been no reported injury or harm to the any patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key6843492
MDR Text Key85118301
Report Number2248146-2017-00370
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeYR
Date Manufacturer Received08/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-